| MEASUREMENT (KEY) |
RANGE |
ACCURACY |
RESOLUTION |
SENSOR LIFE |
SENSOR TYPE |
| Oxygen (O) |
0-21vol% |
0.1 vol% |
0.1vol% |
2 Years |
Electrochemical |
| Carbon Monoxide (C) |
0-4,000 ppm |
± 2% Measured* |
1 ppm |
3 Years |
Electrochemical |
| Carbon Monoxide (V) |
0-40,000 ppm |
± 2% Measured* |
1 ppm |
3 Years |
Electrochemical |
| Nitric Oxide (N) |
0-4,000 ppm |
± 2% Measured* |
1 ppm |
3 Years |
Electrochemical |
| Nitric Oxide (.N) |
0-400 ppm |
± 2% Measured* |
0.1 ppm |
3 Years |
Electrochemical |
| Nitrogen Dioxide (X) |
0-500 ppm |
± 2% Measured* |
1 ppm |
3 Years |
Electrochemical |
| Nitrogen Dioxide (.X) |
0-50 ppm |
± 2% Measured* |
0.1 ppm |
3 Years |
Electrochemical |
| Sulfur Dioxide (S) |
0-5,000 |
± 2% Measured* |
1 ppm |
3 Years |
Electrochemical |
| Combustibles (H) |
0-6.00% vol. |
± 2% Measured* |
0.01% |
5years |
Peellister |
| Gas Temperature |
32°-1800°F |
± 2% Measured* |
1°F |
10 Years |
NiCrNi |
| Ambient Temperature |
0-250°F |
± 2% Measured* |
1 deg F |
10 Years |
Semi-Conductor |
| Draft / Pressure |
± 40" H2o |
± 2% Measured* |
0.1% H2O |
10 Years |
DMS |
| Carbon Dioxide |
0-20% |
±5% Measured* |
0.1% vol% |
4 Years |
Infrared |
| Carbon Monoxide |
1000-63,000ppm |
±5% Measured* |
10ppm |
4 Years |
Infrared |
| O2 Correction |
0-20% Oxygen |
|
|
|
|
| Smoke Scale |
0-9 |
|
|
|
| Carbon Dioxide (CO2) |
0-CO2 max of fuel |
Calculated |
|
|
|
| Efficiency |
0-99.9% |
Calculated |
|
|
|
| Exxess Air (Lambda) |
1- Infinity |
Calculated |
|
|
|
| |
|
|
|
|
|
| *ACCURACY: When calibrated prior to use per ECOM America, Ltd. specifications. |
| |
| PHYSICAL |
EASY
Instrument: 17" x 9.75" x 12"
Weight: 28 lb. complete with sampling system
INDUSTRIAL
Instrument: 20" x 9.75" x 12"
Weight: 29 lb. complete with sampling system
|
 |
| |
| ELECTRICAL |
AC: 110/220V 50/60Hz (User selectable)
Batteries: 4-6 hour battery life, 12V lead-acid
Pump: Flow rate of 3.5+ lpm
Display: Backlit, adjustable contrast & zoom, displays all
parameters simultaneously. |
 |
| |
| OPERATING TEMPERATURE |
Core temperature of the instrument is monitored continuously.
Internal Temperature Compensation software assures accurate
sensor response over the range of 20°F to 104°F. |
 |
| |
| |
|
|
|
|
|
| *Optional probe & sample line lengths are available. |
| |
|
|
|
|
|
|