MEASUREMENT (KEY) RANGE ACCURACY RESOLUTION SENSOR LIFE SENSOR TYPE
Oxygen (O) 0-21vol% 0.1 vol% 0.1vol% 2 Years Electrochemical
Carbon Monoxide (C) 0-4,000 ppm ± 2% Measured* 1 ppm 3 Years Electrochemical
Carbon Monoxide (V) 0-40,000 ppm ± 2% Measured* 1 ppm 3 Years Electrochemical
Nitric Oxide (N) 0-4,000 ppm ± 2% Measured* 1 ppm 3 Years Electrochemical
Nitric Oxide (.N) 0-400 ppm ± 2% Measured* 0.1 ppm 3 Years Electrochemical
Nitrogen Dioxide (X) 0-500 ppm ± 2% Measured* 1 ppm 3 Years Electrochemical
Nitrogen Dioxide (.X) 0-50 ppm ± 2% Measured* 0.1 ppm 3 Years Electrochemical
Sulfur Dioxide (S) 0-5,000 ± 2% Measured* 1 ppm 3 Years Electrochemical
Combustibles (H) 0-6.00% vol. ± 2% Measured* 0.01% 5years Peellister
Gas Temperature 32°-1800°F ± 2% Measured* 1°F 10 Years NiCrNi
Ambient Temperature 0-250°F ± 2% Measured* 1 deg F 10 Years Semi-Conductor
Draft / Pressure ± 40" H2o ± 2% Measured* 0.1% H2O 10 Years DMS
Carbon Dioxide 0-20% ±5% Measured* 0.1% vol% 4 Years Infrared
Carbon Monoxide 1000-63,000ppm ±5% Measured* 10ppm 4 Years Infrared
O2 Correction 0-20% Oxygen        
Smoke Scale 0-9
Carbon Dioxide (CO2) 0-CO2 max of fuel Calculated      
Efficiency 0-99.9% Calculated      
Exxess Air (Lambda) 1- Infinity Calculated      
           
*ACCURACY: When calibrated prior to use per ECOM America, Ltd. specifications.
 
PHYSICAL

EASY
Instrument: 17" x 9.75" x 12"
Weight: 28 lb. complete with sampling system


INDUSTRIAL
Instrument: 20" x 9.75" x 12"
Weight: 29 lb. complete with sampling system

 
ELECTRICAL AC: 110/220V 50/60Hz (User selectable)
Batteries: 4-6 hour battery life, 12V lead-acid
Pump: Flow rate of 3.5+ lpm
Display: Backlit, adjustable contrast & zoom, displays all
parameters simultaneously.
 
OPERATING TEMPERATURE Core temperature of the instrument is monitored continuously.
Internal Temperature Compensation software assures accurate
sensor response over the range of 20°F to 104°F.
 
           
*Optional probe & sample line lengths are available.