ECOM America, Ltd.
1628 Oakbrook Drive
Gainesville, GA 30507

Toll Free 877.326.6411
TEL. 770.532.3280
FAX 770.532.3620

 


J2KN PRO Industrial

We've outdone ourselves (again).


Download Brochure
+ + + + + + +
 

COMPARE THE J2KN PRO SERIES MODELS


STANDARD FEATURES

  • Wireless Handheld Remote Display
  • Up to 9 Gas Sensors (6 Electrochemical and 3 NDIR)
  • Sensor Options: O2, CO, CO High, NO, NO2, SO2, Combustibles, CO2, & Methane
  • Data Logging with Memory Card
  • USB Interface
  • Bluetooth Link
  • On-Board Printer
  • 15ft Teflon Sample Line*
  • 1ft Inconel Probe (up to 1832F/1000C)*
  • CO Purge Pump to Prevent Oversaturation
  • Peltier Cooler Sample Conditioner
  • Peristaltic Pump for Automatic Moisture Removal
  • Heated Sensor Block for Thermal Stability
  • Digital Flow Meter Display
  • Fresh Air Purge Switch
  • Dual Range CO Sensors
  • Self Leak Check
  • Flue Gas, Ambient, & Sensor Temperature Sensors
  • Averaging Tests
  • Mass Emission Calculations
  • CO2, Efficiency, Losses, Excess Air, & O2 Correction Calculations
  • *Other lengths available. Please ask an ECOM sales associate for details.



    OPTIONAL ADD-ONS

  • DAS Compliance Testing Software
  • Calibration Gas Kit
  • Heated Sample Line (15ft & 30ft in stock, custom lengths available)
  • Stack Flow Velocity Probe
  • Soot Test Probe
  • 1ft Flex Tip Probe


  • PARTS LIST    TECHNICAL SPECS

    ECOM J2KN

    The J2KN Pro INDUSTRIAL represents the ultimate design in accurate gas analysis and is unsurpassed in the field of premier portable emissions analyzers. If you have a heavy-duty testing schedule, this is the analyzer for you. In addition to holding up to 6 electrochemical sensors, the J2KN Pro INDUSTRIAL supports an NDIR bench to measure CO2, CO up to 63,000ppm, and CH4. This allows measurement of 9 gases in one compact portable analyzer! When matched with our Advanced DAS Compliance Testing Software, the J2KN Pro INDUSTRIAL offers complete automation of the test procedure so you can get your compliance tests done accurately and efficiently. Advanced features include stack flow velocity, NDIR bench, fresh air purge valve, heated sensor block for thermal stability, on-board heated line power connection, dual range CO sensors for high accuracy in rich and lean conditions, and the ability to set up automatic test cycling for unattended operation.

     
    How does it work?
    The J2KN Pro INDUSTRIAL utilizes electrochemical sensors to measure exhaust emissions, most commonly O2, CO, NO, NO2, and SO2 gases and has an available NDIR bench for measuring CO2, high CO, and CH4. The robust filtering system along with the Peltier gas cooler and 9V high-flow pump ensures a clean sample and high flow rate, which is necessary for accuracy. The sensors have electronic monitoring which compensates for temperature changes in even the harshest testing environments, and the CO sensor has a dedicated fresh air pump which is activated if a maximum concentration limit is exceeded, usually 4000ppm. The robust sample conditioning features, high-speed flow rate, and advanced sensor monitoring make the J2KN Pro INDUSTRIAL analyzer unmatched in accuracy and repeatability
    What is included with the J2KN package?
    - Analyzer with metal chassis built in to aluminum-framed carrying case – Under case storage compartment - Handheld display for complete wireless remote control - 15-foot Teflon sample line - 1-foot Inconel probe - Consumables pack: 10-pack particulate filters, 1 spare NOx/SOx filter tube, 2 in-line filters, 1 extra roll of printer paper, vent filter pack, and 1 refill container of NOx/SOx filter media - Moisture drain tubing – T-10 torx wrench - Ambient temp. sensor stick - Calibration magnet – Antenna for base analyzer – Power cord - 2 case lock keys - Padded shoulder strap - ECOM folder with Operations Manual and Calibration Certificate
    What is the warranty?
    1-year warranty on the analyzer and parts
    2-year warranty on electrochemical sensors
    How does the J2KN compare vs. the competition?
    The handheld remote is the most popular feature of the J2KN Series, but when it comes to accuracy, the sample conditioning system of the J2KN Pro INDUSTRIAL is unmatched in the industry. The high-flow pump allows for a sufficient flow rate, which is crucial for accuracy of the electrochemical and NDIR sensors, especially if the sample has to travel more than 30ft to the analyzer. Another key sample conditioning feature that puts the J2KN Series in a league of its own is the sensor manifold, which is designed for precise flow and gas delivery to each sensor and is heated to provide thermal stability for increased accuracy. With this system, each sensor receives a precise amount of sample which is independent of all other sensors, allowing for a repeatable and accurate measure of each target gas. A few other select J2KN Series features include data logging onto a memory card, ability to perform averaging tests, digital flow reading (no more ball flow meter!), capability to input customer information or site information onto the printout, and optional compliance testing software which include Advanced DAS for engine testers and Boiler Test Software for boiler service technicians. And did we mention the handheld remote!
    What testing applications can J2KN be used for?
    The J2KN Pro INDUSTRIAL analyzer is fully loaded with sensor options and features, which allows it to be used in many different applications around the U.S. and worldwide. These applications include emission testing on stationary engines, generators, compressors, boilers, burners, turbines, heating equipment, pumps, diesel engines, mining equipment (to control DPM), construction equipment, laboratory combustion equipment, alternative fuels research, and many other applications. The J2KN analyzer is used for fuel efficiency testing, combustion tuning, maintenance checks, and emissions compliance reporting. The types of fuel that can be tested on the J2KN Series include natural gas, oil, diesel, coal, wood, biomass, butane, propane, biodiesel, and other alternative fuels. Gasoline emissions may be tested with the J2KN Pro INDUSTRIAL, but this requires the NDIR bench (for CO).
    What does the J2KN measure?
    Gas sensor options include O2, CO, NO, NO2, SO2, Combustibles, CO2, high CO, and CH4. Temperature sensors include flue gas temperature, ambient temperature, and sensor temperature. Draft and pressure are standard measured features. A high-range CO sensor is available for rich burn applications. Also, the NO and NO2 sensors are available in "low NOx" with a resolution of 0.1ppm. Calculations include CO2, excess air, losses, combustion efficiency, O2 correction, and mass emissions in lbs/hour, tons/year, or gr/bhp.
    WHAT'S NEW ON THE
    J2KN PRO INDUSTRIAL?

    • NDIR Bench for Measuring CO2, high CO, and CH4
    • Teflon Sample Line
    • Heated Sensor Block
    • Stack Flow Velocity Probe (Option)
    • Self Leak Check
    • Digital Flow Meter Display