ECOM America, Ltd.
1628 Oakbrook Drive
Gainesville, GA 30507

Toll Free 877.326.6411
TEL. 770.532.3280
FAX 770.532.3620



 

 

MEASUREMENT (KEY) RANGE ACCURACY RESOLUTION SENSOR LIFE SENSOR TYPE
Oxygen (O) 0-21vol% ± 0.3 vol% 0.1 vol% 2 Years Electrochemical
Carbon Monoxide (C) 0-4,000 ppm ± 2% Measured* 1 ppm 3 Years Electrochemical
Carbon Monoxide (V) 0-40,000 ppm ± 2% Measured* 1 ppm 3 Years Electrochemical
Nitric Oxide 0-4,000 ppm ± 2% Measured* 1 ppm 3 Years Electrochemical
Gas Temperature 32-1800 F ± 2% Measured* 1 deg F 10 Years NiCrNi
Ambient Temperature 0-250 F ± 2% Measured* 1 deg F 10 Years Semi-Conductor
Draft / Pressure ± 40" H2o ± 2% Measured* 0.1% H2O 10 Years DMS
O2 Correction 0-20% Oxygen        
NOx Calculated NOx by Programming NO2 factor
Carbon Dioxide (CO2) 0-CO2 max of fuel Calculated      
Efficiency 0-99.9% Calculated      
Exxess Air (Lambda) 1- Infinity Calculated      
Losses 0-99.9% Calculated      
*ACCURACY: When calibrated prior to use per ECOM America, Ltd. specifications.
 
PHYSICAL Instrument: 3" x 10" x 4"
Probe: 10" Length x 3/8" OD Inconel w/ Pistol Grip Handle*
Sample Line: 15' high-temp flex hose w/thermocouple wire*
Weight: 1 lb.
 
ELECTRICAL Batteries: 6-8 hour battery life, 6V lead-acid
Pump: Flow rate of 1.5+ lpm
Display: Backlit, adjustable contrast & zoom, displays all
parameters simultaneously.
 
OPERATING TEMPERATURE Core temperature of the instrument is monitored continuously.
Internal Temperature Compensation software assures accurate
sensor response over the range of 20°F to 104°F.
 
           
*Optional probe & sample line lengths are available.